Comments on “Accuracy of Raman lidar water vapor calibration and its applicability to long-term measurements”

Not Accessible

Your library or personal account may give you access

Abstract

In a recent publication, Leblanc and McDermid [Appl. Opt. , 47, 5592 (2008)] proposed a hybrid calibration technique for Raman water vapor lidar involving a tungsten lamp and radiosondes. Measurements made with the lidar telescope viewing the calibration lamp were used to stabilize the lidar calibration determined by comparison with radiosonde. The technique provided a significantly more stable calibration constant than radiosondes used alone. The technique involves the use of a calib ration lamp in a fixed position in front of the lidar receiver aperture. We examine this configuration and find that such a configuration likely does not properly sample the full lidar system optical efficiency. While the technique is a useful addition to the use of radiosondes alone for lidar calibration, it is important to understand the scenarios under which it will not provide an accurate quantification of system optical efficiency changes. We offer examples of these scenarios. Scanning of the full telescope aperture with the calibration lamp can circumvent most of these limitations. Based on the work done to date, it seems likely that the use of multiple calibration lamps in different fixed positions in front of the telescope may provide sufficient redundancy for long-term calibration needs. Further full-aperture scanning experiments, performed over an extended period of time, are needed to determine a “best practice” for the use of multiple calibration lamps in the hybrid technique.

© 2011 Optical Society of America

Full Article  |  PDF Article
More Like This
Accuracy of Raman lidar water vapor calibration and its applicability to long-term measurements

Thierry Leblanc and I. Stuart McDermid
Appl. Opt. 47(30) 5592-5603 (2008)

Lamp mapping technique for independent determination of the water vapor mixing ratio calibration factor for a Raman lidar system

Demetrius D. Venable, David N. Whiteman, Monique N. Calhoun, Afusat O. Dirisu, Rasheen M. Connell, and Eduardo Landulfo
Appl. Opt. 50(23) 4622-4632 (2011)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Figures (6)

You do not have subscription access to this journal. Figure files are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Equations (1)

You do not have subscription access to this journal. Equations are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All Rights Reserved