Laser power stabilization using optical ac coupling and its quantum and technical limits

Not Accessible

Your library or personal account may give you access

Abstract

We demonstrate an active power stabilization of a Nd:YAG laser employing the optical ac-coupling scheme and derive its fundamental quantum limit. This limit is 3dB better than the one encountered in traditional power stabilization schemes. In our experiment, the optical ac coupling improved the shot-noise-limited sensitivity of the stabilization photodetector by a factor of 11.2. With an independent photodetector, we measured a relative power stability of 3.7×109Hz1/2 at frequencies of around 200kHz. A detailed investigation of the performance limit of our experiment revealed a novel noise source that disturbed the fundamental mode field in the optical resonator. This effect could be of relevance to many precision experiments using optical resonators.

© 2009 Optical Society of America

Full Article  |  PDF Article
More Like This
Solid-state laser intensity stabilization at the 10-8 level

Jameson Rollins, David Ottaway, Michael Zucker, Rainer Weiss, and Richard Abbott
Opt. Lett. 29(16) 1876-1878 (2004)

Stabilized high-power laser system for the gravitational wave detector advanced LIGO

P. Kwee, C. Bogan, K. Danzmann, M. Frede, H. Kim, P. King, J. Pöld, O. Puncken, R. L. Savage, F. Seifert, P. Wessels, L. Winkelmann, and B. Willke
Opt. Express 20(10) 10617-10634 (2012)

Optical ac coupling to overcome limitations in the detection of optical power fluctuations

Patrick Kwee, Benno Willke, and Karsten Danzmann
Opt. Lett. 33(13) 1509-1511 (2008)

Cited By

You do not have subscription access to this journal. Cited by links are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Figures (7)

You do not have subscription access to this journal. Figure files are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Equations (11)

You do not have subscription access to this journal. Equations are available to subscribers only. You may subscribe either as an Optica member, or as an authorized user of your institution.

Contact your librarian or system administrator
or
Login to access Optica Member Subscription

Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All Rights Reserved